Remote control commandsMXO 4 Series
676User Manual 1335.5337.02 ─ 07
For more information, see:
www.rohde-schwarz.com/rc-via-scpi, chapter "Command sequence and synchroni-
zation"
Overview of statistic commands
The remote commands for the power analysis measurements can include statistic
results. In the following the abbreviations after the commands have the following
meaning:
Command Description
POWer<m>:xxx:RESult:nnn:[:ACTual]?
Current measurement result
POWer<m>:xxx:RESult:nnn::AVERage?
Average of the measurement results
POWer<m>:xxx:RESult:nnn:RMS?
RMS value of the measurement results
POWer<m>:xxx:RESult:nnn::PPEak?
Positive peak value of the measurement results
POWer<m>:xxx:RESult:nnn::NPEak?
Negative peak value of the measurement results
POWer<m>:xxx:RESult:xxx::STDDev?
Standard deviation of the measurement results
General settings....................................................................................................676
Power quality.........................................................................................................677
Power harmonics.................................................................................................. 685
Reference level..................................................................................................... 693
Power analysis data..............................................................................................697
17.15.2.1 General settings
POWer<m>[:ENABle]..................................................................................................... 676
POWer<m>:TYPE..........................................................................................................676
POWer<m>:ASKew[:EXECute]........................................................................................677
EXPort:RESult:SELect:POWer<m>..................................................................................677
POWer<m>[:ENABle] <State>
Enables the power measurement instance.
Suffix:
<m>
.
1...3, index of the power analysis measurement
Parameters:
<State> ON | OFF
Usage: Asynchronous command
Manual operation: See "Add" on page 273
POWer<m>:TYPE <Type>
Sets the type for the respective power analysis measurement.