Parameters:
<Index_Operator> EQUal | INRange | RANGe
*RST: INRange
Usage: Asynchronous command
Manual operation: See "Edit" on page 326
17.16.2.3 Trigger
Trigger commands for serial buses affect the current trigger source. Therefore, set the
trigger source to the required serial bus with TRIGger:EVENt<ev>:SOURce before
sending the bus-specific trigger commands.
TRIGger:SPI:TYPE........................................................................................................ 719
TRIGger:SPI:DMINpattern.............................................................................................. 719
TRIGger:SPI:DPOSition..................................................................................................720
TRIGger:SPI:FCONdition................................................................................................720
TRIGger:SPI:PALignment............................................................................................... 720
TRIGger:SPI:TYPE <Type>
Selects the trigger type for SPI analysis.
Parameters:
<Type> FRSTart | FRENd | MOSI | MISO
FRSTart
Triggers on the beginning of the frame.
FRENd
Triggers on the end of the frame.
MOSI
Triggers on a specified data pattern in that is expected on the
MOSI line.
MISO
Triggers on a specified data pattern in that is expected on the
MISO line.
*RST:
FRSTart
Usage: Asynchronous command
Manual operation: See "Type" on page 328
TRIGger:SPI:DMINpattern <DataPattern>
Specifies a data bit pattern, or sets the start value of a pattern range.
Parameters:
<DataPattern>
Usage: Asynchronous command
Manual operation: See "Data" on page 329
Protocols