Remote control commandsMXO 4 Series
730User Manual 1335.5337.02 ─ 07
NACK
Missing acknowledge bit. To localize specific missing acknowl-
edge bits, use:
TRIGger:I2C:ADNack
TRIGger:I2C:DWNack
TRIGger:I2C:DRNack
ADDRess
Triggers on one specific address
DATA
Triggers on a specific data
ADAT
Triggers on a combination of address and data condition.
*RST:
STARt
Usage: Asynchronous command
Manual operation: See "Type" on page 345
TRIGger:I2C:ACCess <RWBitAddress>
Toggles the trigger condition between read and write access of the primary. Select
"Either" if the transfer direction is not relevant for the trigger condition.
Parameters:
<RWBitAddress> READ | WRITe | EITHer
*RST: EITHer
Usage: Asynchronous command
Manual operation: See "R/W bit" on page 347
TRIGger:I2C:ACONdition <AddrOptor>
Sets the operator to set a specific address or an address range. The address values
are set with TRIGger:I2C:ADDRess and TRIGger:I2C:ADDTo.
Parameters:
<AddrOptor> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
*RST: EQUal
Usage: Asynchronous command
TRIGger:I2C:ADDRess <Address>
Triggers on the specified address, or sets the start value of an address range depend-
ing on the condition set with TRIGger:I2C:ACONdition.
Parameters:
<Address> List of comma-separated values