Remote control commandsMXO 4 Series
732User Manual 1335.5337.02 ─ 07
TRIGger:I2C:DMIN <Data>
Specifies the data bit pattern, or sets the start value of a data pattern range. Enter the
bytes in MSB first bit order. The maximum pattern length is 64 bit. Waveform data is
compared with the pattern byte-by-byte.
Parameters:
<Data>
Usage: Asynchronous command
Manual operation: See "Data" on page 348
TRIGger:I2C:DPOSition
<DataPosition>
Sets the number of data bytes to be skipped after the address.
Parameters:
<DataPosition> Range: 1 to 4096
Increment: 1
*RST: 1
Usage: Asynchronous command
Manual operation: See "Position" on page 348
TRIGger:I2C:DRNack <DataReadNack>
Triggers on the end of the read process when the controller reads data from the target.
This NACK is sent according to the protocol definition, it is not an error.
Parameters:
<DataReadNack> ON | OFF
*RST: ON
Usage: Asynchronous command
Manual operation: See "No Ack conditions" on page 347
TRIGger:I2C:DWNack <DataWriteNack>
Triggers if a date acknowledge bit is missing - the addressed target does not accept
the data.
Parameters:
<DataWriteNack> ON | OFF
*RST: ON
Usage: Asynchronous command
Manual operation: See "No Ack conditions" on page 347
Protocols