Remote control commandsMXO 4 Series
761User Manual 1335.5337.02 ─ 07
SBUS<sb>:UART:WORD<w>:STARt?
Returns the start time of the specified word.
Suffix:
<sb>
.
1...4, index of the serial bus
<w> index of the word
Return values:
<WordStart> Range: -1E+26 to 1E+26
Increment: 1E-10
*RST: 0
Default unit: s
Usage: Query only
Asynchronous command
SBUS<sb>:UART:WORD<w>:STOP?
Returns the stop time of the specified word.
Suffix:
<sb>
.
1...4, index of the serial bus
<w> index of the word
Return values:
<WordStop> Range: -1E+26 to 1E+26
Increment: 1E-10
*RST: 0
Default unit: s
Usage: Query only
Asynchronous command
SBUS<sb>:UART:WORD<w>:STATe?
Returns the status of the specified word.
Suffix:
<sb>
.
1...4, index of the serial bus
<w> index of the word
Return values:
<WordState> OK | BREak | STERror | SPERror | PRERror | INComplete
OK: the frame is valid.
BREak: stop bit error with 0x00 word
STERror: start error, incorrect start bit
SPERror: stop error, incorrect stop bit
PRERror: parity error, incorrect parity bit.
INComplete: The frame is not completely contained in the
acquisition. The acquired part of the frame is valid.
Protocols