EasyManua.ls Logo

Rohde & Schwarz MXO 4 Series - Page 769

Rohde & Schwarz MXO 4 Series
925 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commandsMXO 4 Series
769User Manual 1335.5337.02 ─ 07
ID
Identifier: Sets the trigger to one specific identifier or an identifier
range.
To set the identifier, use TRIGger:CAN:ICONdition
,TRIGger:CAN:IMAX, and TRIGger:CAN:IMIN.
IDDT
Identifier and data: Combination of identifier and data conditions.
To set the identifier condition, use TRIGger:CAN:ICONdition,
TRIGger:CAN:IMIN, and TRIGger:CAN:IMAX.
To set the data condition, use TRIGger:CAN:DCONdition and
TRIGger:CAN:DMIN.
ERRC
Error condition: Define the error types with:
TRIGger:CAN:ACKerror
TRIGger:CAN:BITSterror
TRIGger:CAN:CRCerror
TRIGger:CAN:FORMerror
TRIGger:CAN:FDATa:SCERror
*RST:
STOF
Usage: Asynchronous command
Manual operation: See "Type" on page 378
TRIGger:CAN:DCONdition <DataOperator>
Sets the operator to set a specific data pattern or a data pattern range.
Parameters:
<DataOperator> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, not equal, less than, less or equal than, greater than,
greater or equal than. These conditions require one data pattern
to be set with TRIGger:CAN:DMIN.
*RST: EQUal
Usage: Asynchronous command
Manual operation: See "Data pattern" on page 383
TRIGger:CAN:DMIN <DataPattern>
Sets a data pattern, or sets the start value of a data pattern range.
Parameters:
<DataPattern> List of comma separated values
Usage: Asynchronous command
Manual operation: See "Data pattern" on page 383
Protocols

Table of Contents

Related product manuals