Remote control commandsMXO 4 Series
771User Manual 1335.5337.02 ─ 07
ERR: error
OVLD: overload
*RST: CBFF
Usage: Asynchronous command
Manual operation: See "Frame type" on page 380
TRIGger:CAN:ICONdition <IdOperator>
Sets the operator to set a specific identifier or an identifier range.
Parameters:
<IdOperator> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, not equal, less than, less or equal than, greater than,
greater or equal than. These conditions require one identifier
pattern to be set with TRIGger:CAN:IMIN.
INRange | OORange
In range / out of range: Set the minimum and maximum value of
the range with TRIGger:CAN:IMIN and TRIGger:CAN:IMAX.
*RST: EQUal
Usage: Asynchronous command
Manual operation: See "Identifier" on page 382
TRIGger:CAN:IMAX <IdPattern>
Sets the end value of an identifier range if TRIGger:CAN:ICONdition is set to
INRange or OORange.
Parameters:
<IdPattern> List of comma separated values
Usage: Asynchronous command
Manual operation: See "Identifier" on page 382
TRIGger:CAN:IMIN <IdPattern>
Specifies a message identifier pattern, or sets the start value of an identifier range.
Parameters:
<IdPattern> List of comma separated values
Usage: Asynchronous command
Manual operation: See "Identifier" on page 382
Protocols