Remote control commandsMXO 4 Series
772User Manual 1335.5337.02 ─ 07
TRIGger:CAN:ITYPe <IdentifierType>
Selects the format of data and remote frames.
Remote frames are not available in the CAN FD protocol.
Parameters:
<IdentifierType> B11 | B29
B11
11-bit identifier (standard format). The instrument triggers on the
sample point of the IDE bit.
B29
29-bit identifier (extended format). The instrument triggers on the
sample point of the RTR bit.
ANY
The ID type and ID pattern are not relevant for the trigger condi-
tion.
*RST: B11
Usage: Asynchronous command
Manual operation: See "ID type" on page 383
TRIGger:CAN:FORMerror <FormError>
Triggers when a fixed-form bit field contains one or more illegal bits.
Available, if TRIGger:CAN:TYPE is set to ERRC.
Parameters:
<FormError> ON | OFF
*RST: ON
Usage: Asynchronous command
Manual operation: See "Error conditions" on page 384
TRIGger:CAN:FDATa:SCERror <StuffCntErr>
Triggers on stuff count errors. A stuff bit error occurs, if more than five consecutive bits
of the same level occur on the bus.
Available, if TRIGger:CAN:TYPE is set to ERRor.
Parameters:
<StuffCntErr> ON | OFF
*RST: ON
Usage: Asynchronous command
Manual operation: See "Error conditions" on page 384
Protocols