TRIGger:CAN:ACKerror <AckError>
Triggers when the transmitter does not receive an acknowledgment - a dominant bit
during the ACK Slot.
Available, if TRIGger:CAN:TYPE is set to ERRC.
Parameters:
<AckError> ON | OFF
*RST: ON
Usage: Asynchronous command
Manual operation: See "Error conditions" on page 384
TRIGger:CAN:BITSterror <BitStuffError>
Triggers if a stuff error occurs - when the 6th consecutive equal bit level in the men-
tioned fields is detected.
Available, if TRIGger:CAN:TYPE is set to ERRC.
Parameters:
<BitStuffError> ON | OFF
*RST: ON
Usage: Asynchronous command
Manual operation: See "Error conditions" on page 384
TRIGger:CAN:CRCerror <ChecksumError>
Triggers on CRC errors. A CRC error occurs when the CRC calculated by the receiver
differs from the received value in the CRC sequence.
Available, if TRIGger:CAN:TYPE is set to ERRC.
Parameters:
<ChecksumError> ON | OFF
*RST: ON
Usage: Asynchronous command
Manual operation: See "Error conditions" on page 384
TRIGger:CAN:FDATa:BRS <BRS_Bit>
Sets the bit rate switch bit.
Parameters:
<BRS_Bit> ONE | ZERO | DC
ONE: the bit rate switches from the bit rate of the arbitration
phase to the faster data rate.
*RST: ONE