Remote control commandsMXO 4 Series
774User Manual 1335.5337.02 ─ 07
Usage: Asynchronous command
Manual operation: See "BRS, ESI" on page 383
TRIGger:CAN:FDATa:ESI <ESI_Bit>
Sets the error state indicator bit.
Parameters:
<ESI_Bit> ONE | ZERO | DC
DC: do not care, bit is nor relevant.
*RST: DC
Usage: Asynchronous command
Manual operation: See "BRS, ESI" on page 383
TRIGger:CAN:XDATa:AF:CONDition <AfOperator>
Sets the comparison condition for the acceptance field to a specific value or a range.
Parameters:
<AfOperator> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, not equal, less than, less or equal than, greater than,
greater or equal than. These conditions require one pattern to be
set with TRIGger:CAN:XDATa:AF:MIN.
INRange | OORange
In range / out of range: Set the minimum and maximum value of
the range with TRIGger:CAN:XDATa:AF:MIN and TRIGger:
CAN:XDATa:AF:MAX.
*RST: EQUal
Usage: Asynchronous command
Manual operation: See "AF" on page 384
TRIGger:CAN:XDATa:AF:MAX <AfPatternTo>
Sets the end value of an acceptance field if TRIGger:CAN:XDATa:AF:CONDition is
set to INRange or OORange.
Parameters:
<AfPatternTo> List of comma separated values
Usage: Asynchronous command