Remote control commandsMXO 4 Series
797User Manual 1335.5337.02 ─ 07
Parameters:
<Threshold>
Usage: Asynchronous command
Manual operation: See "Threshold" on page 395
SBUS<sb>:LIN:POLarity <Polarity>
Defines the idle state of the bus. The idle state is the recessive state and corresponds
to a logic 1.
Suffix:
<sb>
.
1...4, index of the serial bus
Parameters:
<Polarity> IDLLow | IDLHigh
*RST: IDLHigh
Usage: Asynchronous command
Manual operation: See "Polarity" on page 395
SBUS<sb>:LIN:STANdard <Standard>
Selects the version of the LIN standard.
Suffix:
<sb>
.
1...4, index of the serial bus
Parameters:
<Standard> V1X | V2X | J2602 | AUTO
*RST: AUTO
Usage: Asynchronous command
Manual operation: See "LIN standard" on page 394
SBUS<sb>:LIN:POSition <LIN position>
Sets the vertical position of the LIN signal.
Suffix:
<sb>
.
1...4, index of the serial bus
Parameters:
<LIN position>
Usage: Asynchronous command
SBUS<sb>:LIN:SCALe <LIN scale>
Set the vertical scale of the indicated LIN signal.
Protocols