Remote control commandsMXO 4 Series
811User Manual 1335.5337.02 ─ 07
SBUS<sb>:LIN:FRAMe<fr>:STATus?
Returns the overall state of the selected frame.
Suffix:
<sb>
.
1...4, index of the serial bus
<fr> index of the frame
Return values:
<FrameState> OK | WAKeup | CHCKsum | PRERror | STERror | SYERror |
LNERror | UNK | INComplete
WAKeup: wake-up frame.
CHCKsum: checksum error
PRERror: parity error in identifier
STERror: stop error
SYERror: synchronization error
LNERror: unexpected length
UNK: unknown error
*RST: OK
Usage: Query only
Asynchronous command
SBUS<sb>:LIN:FRAMe<fr>:STOP?
Returns the stop time of the selected frame.
Suffix:
<sb>
.
1...4, index of the serial bus
<fr> index of the frame
Return values:
<FrameStop> Range: -1E+26 to 1E+26
Increment: 1E-10
*RST: 0
Default unit: s
Usage: Query only
Asynchronous command
SBUS<sb>:LIN:FRAMe<fr>:PIDentifier?
Returns the protected identifier of the selected frame.
Suffix:
<sb>
.
1...4, index of the serial bus
<fr> index of the frame