EasyManua.ls Logo

Rohde & Schwarz MXO 4 Series - SPMI (Option R&S MXO4-K550)

Rohde & Schwarz MXO 4 Series
925 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commandsMXO 4 Series
812User Manual 1335.5337.02 ─ 07
Return values:
<Protected identifer> Range: 0 to 255
Increment: 1
*RST: 0
Usage: Query only
Asynchronous command
17.16.7 SPMI (option R&S MXO4-K550)
Before MXO 4 can accept the specific bus commands, make sure that:
The decoding of the specified bus is enabled with SBUS<sb>[:STATe].
The bus is set to the correct type with SBUS<sb>:TYPE.
Configuration.........................................................................................................812
Filter...................................................................................................................... 816
Decode results...................................................................................................... 822
17.16.7.1 Configuration
SBUS<sb>:SPMI:GIDValue.............................................................................................812
SBUS<sb>:SPMI:GSIDenable.........................................................................................813
SBUS<sb>:SPMI:GTCHenable........................................................................................813
SBUS<sb>:SPMI:GTWDith............................................................................................. 813
SBUS<sb>:SPMI:SCLK:HYSTeresis................................................................................ 814
SBUS<sb>:SPMI:SCLK:SOURce.................................................................................... 814
SBUS<sb>:SPMI:SCLK:THReshold.................................................................................814
SBUS<sb>:SPMI:SDATa:HYSTeresis...............................................................................814
SBUS<sb>:SPMI:SDATa:SOURce................................................................................... 815
SBUS<sb>:SPMI:SDATa:THReshold................................................................................815
SBUS<sb>:SPMI:POSition..............................................................................................815
SBUS<sb>:SPMI:SCALe................................................................................................ 815
SBUS<sb>:SPMI:GIDValue <GSID>
Sets a value for the group sub index. Available, if SBUS<sb>:SPMI:GSIDenable is
set to ON.
Suffix:
<sb>
.
1...4, index of the serial bus
Parameters:
<GSID> Range: 0 to 15
Increment: 1
*RST: 0
Usage: Asynchronous command
Manual operation: See "Use GSID, GSID (hex)" on page 407