Remote control commandsMXO 4 Series
820User Manual 1335.5337.02 ─ 07
Usage: Asynchronous command
Manual operation: See "Edit" on page 410
SBUS<sb>:SPMI:FILTer:ERENable <ErrorName>,<Enabler>
SBUS<sb>:SPMI:FILTer:ERRor<n>:ENABle <Enable>
Defines the error type to be filtered on.
Suffix:
<sb>
.
1...4, index of the serial bus
<n> index of the error
Parameters:
<Enable> ON | OFF
*RST: ON
Usage: Asynchronous command
Manual operation: See "Error type" on page 411
SBUS<sb>:SPMI:FILTer:IOPerator <Frame>,<Field>,<Operator>
SBUS<sb>:SPMI:FILTer:FRAMe<fr>:FLD<fl>:IOPerator <Index_Operator>
Sets the operator for the index in the selected field of the selected frame.
Suffix:
<sb>
.
1...4, index of the serial bus
<fr> index of the frame
<fl> index of the field
Parameters:
<Index_Operator> EQUal | INRange | RANGe
*RST: INRange
Usage: Asynchronous command
Manual operation: See "Edit" on page 410
SBUS<sb>:SPMI:FILTer:IMIN <Frame>,<Field>,<Data>
SBUS<sb>:SPMI:FILTer:FRAMe<fr>:FLD<fl>:IMIN <Index_Min>
Specifies the index, or sets the start value of an index range.
Suffix:
<sb>
.
1...4, index of the serial bus
<fr> index of the frame
<fl> index of the field
Protocols

Table of Contents