EasyManua.ls Logo

Rohde & Schwarz MXO 4 Series - Page 863

Rohde & Schwarz MXO 4 Series
925 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commandsMXO 4 Series
863User Manual 1335.5337.02 ─ 07
WGENerator<wg>:MODulation:PWM:DCYCle.................................................................. 868
WGENerator<wg>:MODulation:PWM:DEPTh....................................................................869
WGENerator<wg>:MODulation:PWM:FREQuency............................................................ 869
WGENerator<wg>:MODulation:PWM:SYMMetry...............................................................869
WGENerator<wg>:MODulation:PWM[:FUNCtion].............................................................. 870
WGENerator<wg>:MODulation:TYPE.............................................................................. 870
WGENerator<wg>:MODulation[:STATe] <Modulation>
Enables or disables modulation.
Suffix:
<wg>
.
1...2, index of the waveform generator
Parameters:
<Modulation> ON | OFF
*RST: OFF
Usage: Asynchronous command
Manual operation: See "Modulation state" on page 439
WGENerator<wg>:MODulation:AM:DCYCle <SquareDutyCycle>
Sets the duty cycle for a square waveform. The duty cycle expresses for what percent-
age fraction of the period, the waveform is active, i.e. the signal state is high.
Suffix:
<wg>
.
1...2, index of the waveform generator
Parameters:
<SquareDutyCycle> Range: 10 to 90
Increment: 1
*RST: 50
Default unit: %
Usage: Asynchronous command
Manual operation: See "Duty cycle" on page 440
WGENerator<wg>:MODulation:AM:DEPTh <Depth>
Sets the modulation depth, the percentage of the amplitude range that is used for AM
modulation.
Suffix:
<wg>
.
1...2, index of the waveform generator
Parameters:
<Depth> Range: 0 to 100
Increment: 1
*RST: 50
Default unit: %
Waveform generator (option R&S MXO4-B6)