Remote control commandsMXO 4 Series
865User Manual 1335.5337.02 ─ 07
WGENerator<wg>:MODulation:CARRier:FREQuency <FreqCarrierAlias>
Sets the frequency of the carrier signal for a modulation waveform.
Suffix:
<wg>
.
1...2, index of the waveform generator
Parameters:
<FreqCarrierAlias> Range: 0.001 to 100000000
Increment: 0.001
*RST: 1000000
Default unit: Hz
Usage: Asynchronous command
WGENerator<wg>:MODulation:CARRier:PERiod <PeriodCarrierAlias>
Sets the period of the carrier signal for a modulation waveform.
Suffix:
<wg>
.
1...2, index of the waveform generator
Parameters:
<PeriodCarrierAlias> Range: 1E-08 to 1000
Increment: 1
*RST: 1E-06
Default unit: s
Usage: Asynchronous command
WGENerator<wg>:MODulation:FM:DCYCle <SquareDutyCycle>
Sets the duty cycle for a square waveform. The duty cycle expresses for what percent-
age fraction of the period, the waveform is active, i.e. the signal state is high.
Suffix:
<wg>
.
1...2, index of the waveform generator
Parameters:
<SquareDutyCycle> Range: 10 to 90
Increment: 1
*RST: 50
Default unit: %
Usage: Asynchronous command
Manual operation: See "Duty cycle" on page 440
WGENerator<wg>:MODulation:FM:DEViation <Deviation>
Sets the frequency deviation, the maximum difference between the FM modulated sig-
nal and the carrier signal.
Waveform generator (option R&S
MXO4-B6)