Parameters:
<LevelDC> Range: 0 to 10
Increment: 0.01
*RST: 0
Default unit: Vpp
Usage: Asynchronous command
Manual operation: See "Noise level for DC" on page 438
WGENerator<wg>:MODulation:NLABsolute?
Queries the level of the noise in volts.
Suffix:
<wg>
.
1...2, index of the waveform generator
Return values:
<LevAbs> Range: 0 to 12
Increment: 0.1
*RST: 0
Default unit: Vpp
Usage: Query only
Asynchronous command
WGENerator<wg>:MODulation:NLPCent <LevelPct>
Sets the level of the noise in percentage of the set "Amplitude" output of the signal.
Suffix:
<wg>
.
1...2, index of the waveform generator
Parameters:
<LevelPct> Range: 0 to 100
Increment: 0.1
*RST: 0
Default unit: %
Usage: Asynchronous command
Manual operation: See "Noise level in %" on page 437
WGENerator<wg>:MODulation:PWM:DCYCle <SquareDutyCycle>
Sets the duty cycle for a square waveform. The duty cycle expresses for what percent-
age fraction of the period, the waveform is active, i.e. the signal state is high.
Suffix:
<wg>
.
1...2, index of the waveform generator