Remote control commandsMXO 4 Series
869User Manual 1335.5337.02 ─ 07
Parameters:
<SquareDutyCycle> Range: 10 to 90
Increment: 1
*RST: 50
Default unit: %
Usage: Asynchronous command
Manual operation: See "Duty cycle" on page 440
WGENerator<wg>:MODulation:PWM:DEPTh <Depth>
Sets the modulation depth, the percentage of the amplitude range that is used for
PWM modulation.
Suffix:
<wg>
.
1...2, index of the waveform generator
Parameters:
<Depth> Range: 0 to 100
Increment: 1
*RST: 50
Default unit: %
Usage: Asynchronous command
Manual operation: See "Modulation depth" on page 440
WGENerator<wg>:MODulation:PWM:FREQuency <Frequency>
Sets the frequency of the modulating waveform for PWM modulation.
Suffix:
<wg>
.
1...2, index of the waveform generator
Parameters:
<Frequency> Range: 0.001 to 1000000
Increment: 0.001
*RST: 1000
Default unit: Hz
Usage: Asynchronous command
Manual operation: See "Frequency" on page 440
WGENerator<wg>:MODulation:PWM:SYMMetry <RampSymmetry>
Sets the symmetry for the PWM ramp modulation waveform, the percentage of time
that the waveform is rising.
Suffix:
<wg>
.
1...2, index of the waveform generator
Waveform generator (option R&S MXO4-B6)