Remote control commandsMXO 4 Series
874User Manual 1335.5337.02 ─ 07
WGENerator<wg>:SWEep:TIME <Time>
Sets the duration of the sweep.
Suffix:
<wg>
.
1...2, index of the waveform generator
Parameters:
<Time> Range: 0.001 to 500
Increment: 1
*RST: 0.001
Default unit: s
Usage: Asynchronous command
Manual operation: See "Sweep time" on page 442
WGENerator<wg>:SWEep:FEND
<StopFrequency>
Sets the stop frequency of the sweep signal.
Suffix:
<wg>
.
1...2, index of the waveform generator
Parameters:
<StopFrequency> Range: 0.001 to 100000000
Increment: 0.001
*RST: 1000000
Default unit: Hz
Usage: Asynchronous command
Manual operation: See "Stop frequency" on page 442
17.18.2 Synchronize settings
WGENerator<wg>:COUPling:STATe................................................................................ 874
WGENerator<wg>:COUPling:AMPLitude..........................................................................875
WGENerator<wg>:COUPling:PHASeshift......................................................................... 875
WGENerator<wg>:COUPling[:FREQuency]...................................................................... 875
GENerator:SYNC[:COMBination].....................................................................................876
WGENerator<wg>:COUPling:STATe <CoupleEnable>
Enables the coupling of the generators, with the selected set of parameters: amplitude
and frequency.
Suffix:
<wg>
.
1..2
Specifies the prime generator.
Waveform generator (option R&S MXO4-B6)