EasyManua.ls Logo

Rohde & Schwarz MXO 4 Series - Status:operation Register

Rohde & Schwarz MXO 4 Series
925 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commandsMXO 4 Series
877User Manual 1335.5337.02 ─ 07
Usage: Setting only
SCPI confirmed
17.19.2 STATus:OPERation register
STATus:OPERation commands provide information on the activity of the instrument.
See also: Chapter 16.5.3.3, "STATus:OPERation register", on page 467.
STATus:OPERation:CONDition?......................................................................................877
STATus:OPERation[:EVENt]?..........................................................................................877
STATus:OPERation:ENABle............................................................................................877
STATus:OPERation:CONDition?
STATus:OPERation[:EVENt]?
The CONDition command returns information on actions the instrument is currently
executing. The contents of the register is retained.
The EVENt command returns information on actions the instrument has executed
since the last reading. Reading the EVENt register deletes its contents.
Bits:
0 = ALIGnment
2 = AUToset
4= MEASuring
5= WTRIgger (wait for trigger)
6= TRIggered
Example:
For an example, see Example"Querying with STATus:OPERa-
tion:CONDition" on page 885.
Usage: Query only
STATus:OPERation:ENABle <Enable>
Controls the ENABle part of the STATus:OPERation register. The ENABle defines
which events in the EVENt part of the status register are forwarded to the OPERation
summary bit (bit 7) of the status byte. The status byte can be used to create a service
request.
Parameters:
<Enable> Range: 1 to 65535
Increment: 1
Example:
STATus:OPERation:ENABle 5
The ALIGnment event (bit 0) and AUToset event (bit 2) are for-
warded to the OPERation summary bit of the status byte.
Status reporting

Table of Contents