Remote Control - Commands R&S NRP-Z11/21/31/41/61/211/221
1137.7470.12 6.18 E-7
T
START
T
STOP
Time
P
o
w
e
r
Trigger level
Triggering
(Start)
Triggering
(Stop)
Measurement
window
Fig. 6-4 Effect of SENSe:TIMing:EXCLude:STARt and :STOP in the Burst Average mode
T
START
T
STOP
Time
P
o
w
e
r
Nominal Width Nominal Width
Trigger Delay (< 0) Triggering
(Start)
Measurement
window
Fig. 6-5 Effect of SENSe:TIMing:EXCLude:STARt and :STOP in the Timeslot Average mode
SENSe:TRACe:AVERage:COUNt[?] 1 to 8192
SENSe:TRACe:AVERage:COUNt is used to set the averaging factor for the Trace mode. To put the
setting into effect, you must also activate averaging for the Trace mode and deactivate the Realtime
mode (settings SENSe:TRACe:AVERage:STATe ON and SENSe:TRACe:REALtime OFF). When trace
averaging is activated, the measurement is always performed with chopper stabilization, i.e. the number
of recorded sampling sequences is twice as large as the averaging factor (alternating measurements
with reversed and non-reversed polarity). The result is that two sampling sequences are recorded if the
averaging factor is 1, eight if the averaging factor is 4, and so forth. If the averaging factor transmitted to
the sensor is not a power of two, the parameter is rounded up or off to the next power-of-two number.