R&S NRP-Z11/21/31/41/61/211/221 Remote Control - Commands
1137.7470.12 6.25 E-7
TEST
Table 6-6 Commands of the TEST system
Command Parameter Unit Remarks
TEST:SENSor? Query only
TEST:SENSor?
TEST:SENSor? triggers a selftest of the sensor. In contrast to *TST, this command yields detailed
information, which is useful for troubleshooting.
No signal may be applied to the sensor while the selftest is running.
If the selftest is carried
out with a signal being present, error messages may erroneously
be output for the test steps "Offset Voltages" and/or "Noise Voltages".