R&S NRP-Z5x Remote Control - Commands
1137.0463.12 6.3 E-7
SCPI Commands
The sensor R&S NRP-Z5x is controlled via the groups of commands
CALibration (zeroing)
SENSe (measurement configurations)
SYSTem
TRIGger
SERVice.
CALibration
Table 6-1 Commands of the CALibration system
Command Parameter Unit Remarks
CALibration
:DATA[?] <calibration data set as definite length block>
:LENGth? Bytes Query only
:TEST? Query only
:DEViation? Query only
:REFerence? Query only
:ZERO
:AUTO[?] OFF | ON | ONCE
CALibration:DATA[?] <calibration data set as definite length block>
CALibration:DATA is used for writing a calibration data set in the flash memory of the sensor.
The query yields the calibration data set currently stored in the flash memory as a definite length block.
CALibration:DATA:LENGth?
CALibration:DATA:LENGth? yields the length in bytes of the calibration data set currently stored in the
flash memory. Programs that read out the calibration data set can use this information to determine the
capacity of the buffer memory required.
CALibration:TEST?
The thermal sensors are equipped with an additional test heater on the thermoelectric transducer that
can be fed from a stable DC voltage source inside the sensor. This allows users to test the functionality
and stability of the entire signal processing chain.