R&S NRP-Z5x Remote Control - Commands
1137.0463.12 6.13 E-7
Fig. 6-1 Correction of interactions between the power sensor and the signal source
If the gamma correction is made in connection with an s-parameter correction (setting
SENSe:CORRection:SPDevice:STATe ON), the interaction of the signal source with the s-parameter
device on the one hand and the input of the power sensor on the other hand (depending on the
magnitude of the term | s
12
s
21
|) is corrected. The interaction between the complex reflection coefficient
I
sensor
of the power sensor and the parameter s
22
is always taken into account when the s-parameter
correction is activated – independent of the gamma correction in connection with the signal source.
Fig. 6-2 Correction of interactions between the power sensor, the signal source, and the s-
parameter device
The query yields
• 1 for OFF,
• 2 for ON.
Default setting: OFF
SENSe:SGAMma:MAGNitude[?] 0.0 to 1.0
SENSe:SGAMma:MAGNitude defines the magnitude of the complex reflection coefficient of the signal
source. A value of 0.0 corresponds to ideal match and a value of 1.0 to total reflection.
The query yields the set magnitude.
Default setting: 0.0
SENSe:SGAMma:PHASe[?] –360.0 to 360.0
SENSe:SGAMma:MAGNitude defines the phase angle (in degrees) of the complex reflection coefficient
of the signal source.
The query yields the set phase angle.
Default setting: 0.0 [°]