Remote Control - Commands R&S NRP-Z5x
1137.0463.12 6.22 E-7
Lower limit x and upper limit y of the trigger threshold
SENSe:CORRection:OFFSet:STATe OFF:
x = <lower test limit in W> × 25
y = <upper test limit in W>
SENSe:CORRection:OFFSet:STATe ON:
x = <lower test limit in W> × 25 × 10
OFFSET / 10 dB
y = <upper test limit in W> × 10
OFFSET / 10 dB
<lower test limit in W>: The nominal lower test limit of the sensor or of the twoport/sensor
combination (with activated s-parameter correction)
<upper test limit in W>: The nominal upper test limit of the sensor or of the twoport/sensor
combination (with activated s-parameter correction)
The nominal test limits can be read out with the SYSTem:INFO? "MINPOWER" and SYSTem:INFO?
"MAXPOWER" command.
Default setting: 10 × x
TRIGger:SLOPe[?] POSitive | NEGative
TRIGger:SLOPe defines the edge of the trigger event with internal and external triggering. In this
connection, positive means increasing envelope power (with internal triggering) or increasing voltage
(with external triggering). This command has no effect in conjunction with trigger sources BUS, HOLD
and IMMediate.
The query yields
• 1 for POSitive,
• 2 for NEGative.
Default setting: POSitive
TRIGger:SOURce[?] HOLD | IMMediate | INTernal | BUS | EXTernal |
TRIGger:SOURce sets the trigger source.
• HOLD: Triggering only with command TRIGger:IMMediate.
• IMMediate: Automatic triggering without explicit event.
• INTernal: Triggering by the measurement signal. Relevant parameters:
TRIGger:LEVel, TRIGger:SLOPe, TRIGger:DELay, TRIGger:HYSTeresis,
TRIGger:HOLDoff.
• BUS: Triggering with command *TRG or TRIGger:IMMediate.
• EXTernal: Triggering via the hardware trigger bus, e.g. via the R&S NRP-Z3 USB adapter
or the R&S NRP base unit. Relevant parameters:
TRIGger:SLOPe, TRIGger:DELay, TRIGger:HOLDoff.
The query yields
• 1 for HOLD,
• 2 for IMMediate,
• 4 for INTernal,
• 8 for BUS,
• 16 for EXTernal.
Default setting: IMMediate