Instrument Functions
NRP2
97User Manual 1173.9140.02 ─ 07
Left
Enter the left end-of-scale value.
Remote command:
DISPlay[:WINDow]<[1]..4>:
METer:LOWer on page 243
Right
Enter the right end-of-scale value.
Remote command:
DISPlay[:WINDow]<[1]..4>:METer:UPPer on page 244
Scrolling between timeslots
In the "Timeslot Av" mode, the timeslot to be displayed can be chosen.
Timeslot
Selects one timeslot out of a complete structure to show its mean power in the active
window.
If you want to display further timeslots simultaneously, a separate window must be
opened and configured for each additional timeslot.
Remote command:
DISPlay[:WINDow]<[1]..4>:
TSLot on page 248
4.4 Configuring measurements
The R&S NRP2 can perform up to four measurements simultaneously. Each measure-
ment is configured separately in a window. Measurement parameters are set in the
"Measurement" menu.
For every window opened via the "Windows" menu, a measurement in the "Measure-
ment" menu can be configured. The measurement channels (A, B, C and D), the pro-
cessing of the measured values (ratio, difference), the result unit and additional infor-
mation (measurement uncertainty, max, min) are defined.
Configuring a measurement does not change the sensor operating modes that have
been set in the "Mode" menu.
In remote control mode, the instrument allows the configuration of a total of 8 measure-
ments. However, only measurement 1 to 4 can be assigned a result window. Measure-
ments number 5 to 8 can only be evaluated via remote control.
4.4.1 The Measurement menu
The "Measurement" menu contains the functions required to configure measurements.
Configuring measurements