Instrument Functions
NRP2
112User Manual 1173.9140.02 ─ 07
●
A warning tone is output (optional, only when "Warning Beep"
is activated, see
below)
●
A signal is issued on analog output "OUT1" (optional, chapter 4.7.2, "Setting the
analog outputs and the trigger input", on page 151).
Fig. 4-43: Limits dialog box
xx.xxxx dBm / Upper
Sets the upper limit value and turns monitoring on/off.
Remote command:
CALCulate<[1]..8>:LIMit:
UPPer[:DATA] on page 230
CALCulate<[1]..8>:LIMit:UPPer:STATe on page 230
xx.xxxx dBm / Lower
Sets the lower limit value and turns monitoring on/off.
Remote command:
CALCulate<[1]..8>:LIMit:LOWer[:DATA] on page 228
CALCulate<[1]..8>:LIMit:LOWer:STATe on page 229
Warning Beep
Activates an acoustic alarm which is output if a limit has been exceeded.
Remote command:
CALCulate<[1]..8>:LIMit:BEEP on page 227
4.4.6 Resetting auxiliary information
For each window, the R&S NRP2 continuously saves data that is required for display-
ing the extreme values and statistical values. These values can be reset any time.
To reset the auxiliary values press the lowest softkey in the "Measurement" menu.
Aux Val Reset
Restarts the search for extreme values and statistical evaluation (if activated in the
"Auxiliaries" dialog box).
Remote command:
CALCulate<[1]..8>:EXTRemes:RESet on page 226
Configuring measurements