Instrument Functions
NRP2
134User Manual 1173.9140.02 ─ 07
4.5.7.3 T'Slot diagram settings
► Configure timeslot measurement in the
"Timeslot Config" dialog box which is
opened via the Window→Graph→"Display"→"T'slot" control panel ("Trace&Statis-
tics" sensor mode).
Fig. 4-60: Timeslot Config dialog box
No of Timeslots
Defines the number of subsequent timeslots within a frame, i.e. a period of the signal.
Remote command:
[SENSe<[1]..4>][:POWer]:
TSLot:COUNt on page 277
Nominal Width
Defines the length of a timeslot.
Example:
If measurements are performed on TDMA signals (GSM/EDGE, PDC, NADC, PHS
etc), the nominal timeslot length must be entered here, i.e. the frame length divided by
the number of timeslots.
Table 4-10: GSM signal
Frame length: 4.615 ms
No. timeslots: 8
⇒ Nominal width: 4.615 / 8 ms = 576.875 μs
Remote command:
[SENSe<[1]..4>][:POWer]:TSLot:
[AVG]:WIDTh on page 278
Excluded from Start / End
Defines the intervals at the start or end of the timeslot which are excluded from the
measurement. Measurement can thus be restricted to the relevant part of a timeslot.
Remote command:
[SENSe<[1]..4>]:
TIMing:EXCLude:
STARt on page 283
[SENSe<[1]..4>]:TIMing:EXCLude:STOP on page 283
Fence
Activates the "Fence" function.
Displaying traces