Remote Control Fundamentals
NRP2
185User Manual 1173.9140.02 ─ 07
The following table lists the respective setting capabilities and the values that are auto-
matically set in manual control.
Table 5-3: Automatic settings for manual selection of a measurement mode
SCPI command Value Remark
Mode Cont Avg
TRIGger[1..4]:SOURce IMMediate
Cannot be modified in manual mode.
TRIGger[1..4]:SLOPe
- Unchanged, is irrelevant.
TRIGger[1..4]:COUNt 1
Cannot be modified in manual mode.
TRIGger[1..4]:DELay
- Unchanged, is irrelevant.
TRIGger[1..4]:HOLDoff 1 NS
Cannot be modified in manual mode.
TRIGger[1..4]:LEVel
- Unchanged, is irrelevant.
[SENSe[1..4]]:TIMing:EXCLude:
STARt
- Unchanged, is irrelevant.
[SENSe[1..4]]:TIMing:EXCLude:STOP
- Unchanged, is irrelevant.
INITiate[1..4]:CONTinuous ON
Cannot be modified in manual mode.
Mode Burst Av
TRIGger[1..4]:SOURce
- Unchanged, is irrelevant.
TRIGger[1..4]:SLOPe
- Unchanged, is irrelevant.
TRIGger[1..4]:COUNt 1
Cannot be modified in manual mode.
TRIGger[1..4]:DELay
- Unchanged, is irrelevant.
TRIGger[1..4]:HOLDoff
Depends on sensor User-selectable.
TRIGger[1..4]:LEVel
Depends on sensor User-selectable.
[SENSe[1..4]]:TIMing:EXCLude:
STARt
Depends on sensor User-selectable.
[SENSe[1..4]]:TIMing:EXCLude:STOP
Depends on sensor User-selectable.
INITiate[1..4]:CONTinuous ON
Cannot be modified in manual mode.
Mode Timegate Av
TRIGger[1..4]:SOURce EXTernal |
INTernal
User-selectable.
TRIGger[1..4]:SLOPe POSitive |
NEGative
User-selectable.
TRIGger[1..4]:COUNt 1
Cannot be modified in manual mode.
TRIGger[1..4]:DELay
Depends on sensor User-selectable.
TRIGger[1..4]:HOLDoff
Depends on sensor User-selectable.
TRIGger[1..4]:LEVel
Depends on sensor User-selectable.
[SENSe[1..4]]:TIMing:EXCLude:
STARt
- Unchanged, is irrelevant.
Automatic Settings for Manual Selection of a Measurement Mode
北京海洋兴业科技股份有限公司(证券代码:839145)