Remote Control - Commands
NRP2
199User Manual 1173.9140.02 ─ 07
Table 6-2: Preset values
Command Preset and *RST value Remark
CALC<[1]..8>:FEED[1..2] "POW:AVER"
The calculate block processes
average power values ("Aver-
age").
CALC<[1]..8>:LIM:BEEP OFF
No audible signal in case of limit
violations.
CALC<[1]..8>:LIM:CLE:AUTO OFF
No automatic reset of the limit
monitoring function when a mea-
surement is started.
CALC<[1]..8>:LIM:FAIL? 0
Limit violation.
CALC<[1]..8>:LIM:FCO? 0
Number of limit violations is reset.
CALC<[1]..8>:LIM:UPP:DATA 0 W or 0 DB
Upper limit value for measure-
ments.
CALC<[1]..8>:LIM:UPP:STAT OFF
No monitoring function for upper
limit value.
CALC<[1]..8>:LIM:LOW:DATA 0 W or 0 DB
Lower limit value for measure-
ments.
CALC<[1]..8>:LIM:LOW:STAT OFF
No monitoring function for lower
limit value.
CALC<[1]..8>:MATH
For command description
see page 231(
The calculate block returns the
measured value of the sensor
assigned to the primary channel.
CALC<[1]..8>:REL 0 DBM or 0 DB
-
CALC<[1]..8>:REL:POW 0 DBM
-
CALC<[1]..8>:REL:RAT 0 DB
-
CALC<[1]..8>:REL:STAT OFF
Measurements are not converted
into a reference value.
DISP:ILL -
Display lighting remains
unchanged.
DISP:MESS OFF
Do not display messages.
DISP:MESS:TEXT -
No change.
DISP:MESS:TYPE MESS
Messages cannot be confirmed
manually.
DISP:SEL 1
First window selected.
DISP:SIZE NORM
Automatic window sizes.
DISP<[1]..4> ON | OFF
A window is opened automatically
for each connected sensor.
DISP<[1]..4>:AVAL NONE
No additional information on mea-
sured values displayed.
DISP<[1]..4>:FORM DIG
Digital result display.