Remote Control - Commands
NRP2
210User Manual 1173.9140.02 ─ 07
With the FETCh and READ commands, the parameter list is used to crosscheck the set-
tings of the instrument with the settings given in the list. If the settings do not agree,
command execution is aborted. FETCh and READ can also be used as low level com-
mands. In this case, they only retrieve current the measurement result (FETCh?) or
start a measurement and retrieve the measurement result (READ?).
CONFigure:...
The Configure:... commands define the calculation function and configure the associ-
ated sensors by means of the parameter list (see chapter 6.4.4, "Parameter list
description", on page 212
for an overview of the parameters). A measurement is not
started. A question mark at the end of the command provides information on the
parameters transferred by the last CONFigure or MEASure command, respectively.
Since the device settings can be changed after a CONFigure, the query does not
return the current device setup. CONFigure is performed internally mostly by SENSe
and CALCulate commands. One or several measurement channels are configured as
follows:
Command Measurement mode:
ContAv Burst Timeslot Trace/Statis-
tics
INIT:CONT OFF OFF OFF OFF
TRIG:SOUR IMM INT EXT EXT
TRIG:COUN 1 1 1 1
TRIG:DEL:AUTO ON ON ON ON
SENS:AVER:COUN:AUTO ON ON ON
-
SENS:AVER:COUN:AUTO:TYPE ON ON ON
-
SENS:AVER:COUN:AUTO:SLOT
- - 1 -
SENS:AVER:STAT ON ON ON ON
SENS:FUNCtion 'POWer:AVG' 'POWer:
BURSt:
AVG'
'POWer:
TSLot:AVG'
'XTIMe:
POWer'
'XPOWer:
CCDF'
'XPOWer:
PDF'
The configuration of the calculation function (CALCulate:MATH
and CALCulate:REL
commands) depends on the given high level command.
FETCh:...?
The FETCh:...? command returns the last valid measurement result. The command
is only completed after a valid measurement has been performed.
High-Level Measurement Commands (CONFigure, MEASure, READ and FETCh)