Parameter lists for the different measurement modes
●
measurement mode ContAV:
[<expected_value> [,<resolution>]] [,<source_list>]
●
measurement mode T'slot:
<tslot_width>, <no_slots>, <start_exclude>, <end_exclude>,
[<expected_value> [,<resolution>]] [,<source_list>]
●
measurement mode Burst:
<dtolerance>, <start_exclude>, <end_exclude>, [<expected_value>
[,<resolution>]] [,<source_list>]
●
measurement mode ContAv (with data buffering)
(<buffered_size>), [<expected_value>[,
<resolution>]]
[,<source_list>]
●
measurement mode Trace:
(<scope_size>),
<capture_time> [,<source_list>]
●
measurement mode Statistics:
(<statistics_size>), <capture_time> [,<source_list>]
The elements of the parameter lists have the following meaning:
Table 6-6: Description of the parameters in the parameter list in alphabetical order
<buffered_size>
Value range: depending on sensor.
Default value: depending on sensor.
Unit: -
Number of requested measured values. Corresponds to command:
[SENSe<[1]..4>][:
POWer][:
AVG]:BUFFer:SIZE on page 272.
<capture_time>
Value range: depending on sensor.
Unit: S
Default value: depending on sensor.
Period within which measured data are captured in the Trace mode. Corre-
sponds to command:
[SENSe<[1]..4>]:TRACe:
TIME on page 293.
Period within which measured data are captured in the STATistics mode. Corre-
sponds to command:
[SENSe<[1]..4>]:STATistics:TIME on page 283.
<dtolerance>
Value range: depending on sensor.
Unit: S
Default value: depending on sensor.
The length of a time interval in which the power level may drop below the trigger
level without the end of a power pulse being detected. Corresponds to com-
mand:
[SENSe<[1]..4>][:
POWer]:
BURSt:DTOLerance on page 274.