Remote Control - Commands
NRP2
237User Manual 1173.9140.02 ─ 07
Unit: s
RST: 0 s
CALCulate<[1]..8>:TRACe:MARKer<[1]..2>:
YPOSition?
Is available with trace measurements. It returns the position of a curve marker on the
power axis if the value is valid. The unit depends on the trace function.
Units: W (for trace function SINGLE) | DB (for trace function RATIO)
Usage: Query only
6.6 CALibration
Suffix [1]..4 denotes the four sensors. Calibration can be performed for only one of
the sensors using header CALibration<[1]..4>:...
or for all sensors using
header CALibration:ALL:....
CALibration<[1]..4>:DATA...............................................................................................237
CALibration<[1]..4>:ZERO:AUTO.....................................................................................237
CALibration:ALL:ZERO:AUTO.........................................................................................237
CALibration<[1]..4>:ZERO:FAST:AUTO........................................................................... 238
CALibration:ALL:ZERO:FAST:AUTO................................................................................238
CALibration<[1]..4>:DATA <data>
Read and write binary calibration data sets for the selected sensor.
Parameters:
<data> block_data
CALibration<[1]..4>:ZERO:AUTO <auto>
CALibration:ALL:ZERO:AUTO <state>
Performs zeroing using the signal at the sensor input(s). The sensor(s) must be dis-
connected from all power sources.
Error messages:
-231 "Data questionable; ZERO ERROR": Signal at the input considerably deviates from 0 W.
-224 "Illegal parameter value": The transfer parameter is a numeric value unequal 0 (corre-
sponds to OFF).
The setting commands accept only the parameters ONCE, LFR and UFR, they start of
zeroing; OFF and ON are ignored.
The queries return the value ON if a calibration is in progress, otherwise value OFF.