Remote Control - Commands
NRP2
244User Manual 1173.9140.02 ─ 07
DISPlay[:WINDow]<[1]..4>:METer:LOWer:POWer <lower_power>
DISPlay[:WINDow]<[1]..4>:ANALog:LOWer:POWer <lower_power>
Sets the lower limit value of the analog scale. This value is used if the unit of the mea-
surement result is Watt, dBm or dBµV, i.e. the measurement result is a power.
Value range: See table 6-12 Value range.
Unit: DBM | W | DBUV
Parameters:
<lower_power> float_value
*RST: -90 DBM
Default unit: DBM
DISPlay[:WINDow]<[1]..4>:METer:LOWer:RATio <lower_ratio>
DISPlay[:WINDow]<[1]..4>:ANALog:LOWer:RATio <lower_ratio>
Sets the lower limit value of the analog scale. This value is used if the unit of the mea-
surement result is dB, percent or the number 1, i.e. the measurement result is a power
ratio.
Value range: See table 6-12 Value range.
Unit: DB | DPCT | O
Parameters:
<lower_ratio> float_value
*RST: -120 DB
Default unit: DB
DISPlay[:WINDow]<[1]..4>:METer:UPPer <upper_limit>
DISPlay[:WINDow]<[1]..4>:ANALog:UPPer <upper_limit>
Sets the upper limit value of the analog scale.
The value range depends on the current output unit of the measured value of the win-
dow.
Unit: See table 6-10 Value ranges, depending on the unit.
Parameters:
<upper_limit> float_value
*RST: 30 DBM or 60 DB
DISPlay[:WINDow]<[1]..4>:METer:UPPer:POWer <upper_limit>
DISPlay[:WINDow]<[1]..4>:ANALog:UPPer:POWer <upper_limit>
Sets the upper limit value of the analog scale. This value is used if the unit of the mea-
surement result is Watt, dBm or dBµV, i.e. the measurement result is a power.
Value range: See table 6-10 Value ranges, depending on the unit.
Unit: DBM | W | DBUV