Remote Control - Commands
NRP2
246User Manual 1173.9140.02 ─ 07
This command corresponds to the <resolution> option of the high-level measurement
commands CONFigure, FETCH, READ and MEASure
.
Note:
Automatic filter ranging must be switched on
(SENSe<[1]..4>:AVERage:COUNt:AUTO ON). DISP<[1]..8>:RES has an effect
on the setting SENS<[1]..4>:AVER:COUN:AUTO:RES of the sensors used in the
measurement function (CALC<[1]..8>:MATH). However, the resolution setting of the
sensors does not affect the resolution of values in the display.
Error messages:
-127 "Invalid numeric data": An invalid number was specified for the resolution.
Parameters:
<resolution> 1 | 0.1 | 0.01 | 0.001
lin = Linear scale of measurement result
log = Logarithmic scale of measurement result
1
lin = 1 significant place
log = No decimal place
0.1
lin = 2 significant places
log = 1 decimal place
0.01
lin = 3 significant places
log = 2 decimal places
0.001
lin = 4 significant places
log = 3 decimal places
*
RST: 0.01
Example:
DISP:WIND1:
RES 0.1
With a measured value of 123.456789 W, the first two places
are free of noise and thus 120 W is displayed.
DISPlay[:WINDow]:SELect <select_window>
Selects a window.
Parameters:
<select_window> 1 | 2 | 3 | 4
*
RST: 1
DISPlay[:WINDow]:SIZE <window_size>
Sets the window size.
DISPlay