Remote Control - Commands
NRP2
258User Manual 1173.9140.02 ─ 07
OUTPut:RECorder<[1]..2>[:STATe] <state>
Switches the analog outputs (OUT1 and OUT2/TRIG on rear panel) ON or OFF. The
magnitude of the output voltage (0 V to 3.3 V) depends on the measured power of the
assigned calculate block (OUTP:REC:FEED
) and the configuration of the characteristic
(OUTP:REC:LIM). As in manual operation, this setting is coupled to other operating
modes of the outputs. Only one mode can be active at any one time:
Connector Analog output
(OUTP:REC
)
TTL output
(OUTP:TTL)
Off Ext. Trigger In
OUT1 x x x -
OUT2/TRIG x - - x
Parameters:
<state> ON | OFF
*RST: output 1: ON; output 2: OFF
OUTPut:ROSCillator[:STATe] <state>
The command switches the optional RF generator (ON) or (OFF).
Parameters:
<state> ON | OFF
*RST: OFF
OUTPut:TTL:ACTive <string>
Specifies whether the value set via OUTP:TTL:LVOL
(LOW) or via OUTP:TTL:HVOL
(HIGH) should be applied if the measured value causes a limit violation at the TTL
output.
Parameters:
<string> LOW | HIGH
*RST: LOW
OUTPut:TTL:FAIL <mode>
Specifies the value which should be applied if the measured value causes a limit viola-
tion at the TTL output.
Note: The OUTP:TTL:ACT
and OUTP:TTL:FAIL commands are identical, except for
the opposite meaning of the parameters.