Remote Control - Commands
NRP2
259User Manual 1173.9140.02 ─ 07
Parameters:
<mode> HIGH
The value set via OUTP:
TTL:LVOL
is applied if the measured
value causes a limit violation at the TTL output.
LOW
The value set via OUTP:TTL:HVOL is applied if the measured
value causes a limit violation at the TTL output.
*RST: HIGH
OUTPut:TTL:FEED <calculate_block>
Sets the calculate block whose limit handling controls the TTL voltage. If the measured
value of this calculate block violates a limit and limit monitoring is active, the voltage
specified in table 6-14 Voltage at TTL output (first line) is output.
Parameters:
<calculate_block> string
Range: "CALC1:LIM","CALC2:LIM","CALC3:LIM","CALC4:L
IM"
*RST: "CALC1:LIM"
OUTPut:TTL:HVOLtage <high_voltage>
Sets the high voltage for the TTL output.
Parameters:
<high_voltage> float_value
Range: 0 V to 3.3 V
*RST: 3.3 V
Default unit:
V
OUTPut:TTL:LVOLtage <low_voltage>
Sets the low voltage for the TTL output.
Note: The value for the lower TTL level may be greater than the value for the upper
TTL level.
Parameters:
<low_voltage> float_value
Range: 0 V to 3.3 V
*
RST: 0 V
Default unit:
V
OUTPut