Remote Control - Commands
NRP2
260User Manual 1173.9140.02 ─ 07
OUTPut:TTL[:STATe] <state>
Activates/deactivates the TTL operating mode of the OUT1 connector (rear panel). The
magnitude of the two possible output voltages can range from 0 V to 3.3 V
(OUTP:TTL:HVOL
and OUTP:TTL:LVOL). Refer to OUTPut:TTL[:STATe]
on page 260 and figure 6-9 Voltage at TTL output to determine which voltage is availa-
ble at the output.
Table 6-14: Voltage at TTL output
OUTP:TTL:ACT LOW
(OUTP:TTL:FAIL HIGH)
OUTP:TTL:ACT HIGH
(OUTP:TTL:FAIL LOW)
Measured value violates limit
OUTP:TTL:LVOL OUTP:TTL:HVOL
Measured value is within limit
OUTP:TTL:HVOL OUTP:TTL:LVOL
Power
Voltage
OUTP:TTL:HVOLT
DISP:LIM:LOW DISP:LIM:UPP
OUTP:TTL:LVOLT
DISP:LIM:LOW DISP:LIM:UPP
OUTP:TTL:ACT HIGH
(OUTP:TTL:FAIL LOW)
OUTP:TTL:ACT LOW
(OUTP:TTL:FAIL HIGH)
Fig. 6-9: Voltage at TTL output
Parameters:
<state> ON | OFF
*RST: OFF
OUTPut:TRIGger[:STATe]
Switches the trigger output ON or OFF. The source of signals on the trigger bus can
either be the (external) trigger input or a sensor in trigger master mode. The output
needs to be switched to OFF before it can be switched ON again. Therefore, it is not
possible to switch directly from OUTP:REC ON to OUTP:TRIG ON.
6.11 SENSe (Sensor Commands)
The commands of the SENSe group configure the power sensors with the exception of
the trigger system. This includes the measurement modes (ContAv, Burst, Timeslot or
Trace), the offset corrections and the filter settings.
Suffix [1]..4 denotes the sensor channel.
SENSe (Sensor Commands)