Remote Control - Commands
NRP2
282User Manual 1173.9140.02 ─ 07
[SENSe<[1]..4>]:STATistics:SCALe:X:POINts <number>
Determines the number of samples on the power axis for the Statistics modes or
returns the current setting. Due to the limited storage capacity of the R&S NRP2, the
number is limited to 1024.
Parameters:
<number> int_value
Range: 1 ... 1024
[SENSe<[1]..4>]:STATistics:SCALe:X:RANGe <range>
Defines, together with the SENS:STAT:SCAL:X:RLEV command, the range on the
power axis over which the statistics function (CCDF/PDF) is calculated (see fig-
ure 6-14 Setting the power axis for statistics measurements).
Parameters:
<range> float_value
Range:
depending on sensor
*RST: depending on sensor
Default unit:
DB
[SENSe<[1]..4>]:STATistics:SCALe:X:RLEVel <rlevel>
Defines, together with the SENS:STAT:SCAL:X:RANG command, the range on the
power axis over which the statistics function (CCDF/PDF) is calculated (see figure
below).
Power
Density/Distribution
SENS:STAT:SCAL:X:RANG
SENS:STAT:SCAL:X:RLEV
Fig. 6-14: Setting the power axis for statistics measurements
Unit: DBM | W | DBUV
SENSe (Sensor Commands)