Remote Control - Commands
NRP2
284User Manual 1173.9140.02 ─ 07
Parameters:
<exclude_stop> float_value
Range: depending on sensor
*
RST: depending on sensor
[SENSe<[1]..4>]:TRACe:
AVERage[:STATe] <state>
Switches the filter function of a sensor on or off for the Trace mode. When the filter is
switched on, the number of measured values set with
[SENSe<[1]..4>]:AVERage:COUNt is averaged. This reduces the effect of noise
so that more reliable results are obtained.
Parameters:
<state> ON |
OFF
*RST: depending on sensor
[SENSe<[1]..4>]:TRACe:AVERage:COUNt <filter_length>
Sets the length of the filter for the Trace mode. The wider the filter, the lower the noise
and the longer it takes to obtain a measured value.
Note:
The SENS:TRAC:AVER:COUN command switches the automatic filter length switch-
over off (SENS:TRAC:AVER:COUN:AUTO OFF).
Parameters:
<filter_length>
The filter length can only be 1, 2, 4, 8, 16,..., 2
n
. If the entered
filter length is not a power of two, the value is rounded to the
next 2
n
value without an error message.
Range: depending on sensor
*
RST: depending on sensor
[SENSe<[1]..4>]:TRACe:AVERage:COUNt:AUTO <state>
Can be used to automatically determine a value for
[SENSe<[1]..4>]:AVERage:COUNt for the trace mode.
SENSe (Sensor Commands)