EasyManua.ls Logo

Rohde & Schwarz NRP2 - Page 288

Rohde & Schwarz NRP2
419 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control - Commands
NRP2
285User Manual 1173.9140.02 ─ 07
Parameters:
<state> ON | OFF | ONCE
ON
The automatic switchover is activated, the sensor always
defines a suitable filter length which can also be queried via
SENS<[1]..4>:
AVER:
COUN?
OFF
The automatic switchover is deactivated.
ONCE
The automatic switchover is deactivated and the suitable filter
length is then defined and set automatically. The setting can
then be queried via SENS<[1]..4>:AVER:COUN?
*RST: depending on sensor
[SENSe<[1]..4>]:TRACe:AVERage:COUNt:AUTO:MTIMe <mtime>
If the R&S NRP2 has to determine the filter length automatically in the Trace mode, the
filter length can become large and thus also the time to fill the filter. An upper time limit
(maximum time) can be set via SENS<[1]..4>:AVER:COUN:
AUTO:MTIME. It should
never be exceeded. Undesired long measurement times can thus be prevented (in the
SENS:AVER:COUN:AUTO:TYPE NSR mode) if the automatic filter length switchover is
on.
Parameters:
<mtime> float_value
Range: depending on sensor
*RST: depending on sensor
Default unit:
S
[SENSe<[1]..4>]:TRACe:AVERage:COUNt:AUTO:RESolution <resolution>
Defines the number of significant places for the Trace mode for linear units and the
number of decimal places for logarithmic units which should be free of noise in the
measurement result. This setting is also performed by the command DISPlay[:
WINDow]<[1]..4>:RESolution on page
245 which tries to set the sensors
involved in the measurement result accordingly. For further details, see chapter 4,
"Instrument Functions", on page 61.
However, SENS<[1]..4>:AVER:COUN:AUTO:RES does not affect the DISPlay
command. The parameters of the two commands are different but have the same
meaning.
SENS<[1]..4>:AVER:COUN:AUTO:RES
1 2 3 4
DISP<[1]..4>:RES
1 0.1 0.01 0.001
SENSe (Sensor Commands)
北京海洋兴业科技股份有限公司(证券代码:839145)
电话:010-62176775
网址:www.hyxyyq.com

Table of Contents

Related product manuals