EasyManua.ls Logo

Rohde & Schwarz NRP2 - Page 292

Rohde & Schwarz NRP2
419 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control - Commands
NRP2
289User Manual 1173.9140.02 ─ 07
[SENSe<[1]..4>]:TRACe:MEASurement:POWer:MAX?
Returns the measured maximum power in the current trace.
Unit: WATT
Usage: Query only
[SENSe<[1]..4>]:TRACe:MEASurement:POWer:MIN?
Returns the measured minimum power in the current trace.
Unit: WATT
Usage: Query only
[SENSe<[1]..4>]:TRACe:MEASurement:POWer:PULSe:BASE?
Returns the base power, computed in the pulse level analysis histogram of the current
trace data.
Unit: WATT
Usage: Query only
[SENSe<[1]..4>]:TRACe:MEASurement:POWer:PULSe:TOP?
Returns the power of the first detected complete pulse of the current trace data, includ-
ing rising and falling edges.
Unit: WATT
Usage: Query only
[SENSe<[1]..4>]:TRACe:MEASurement:PULSe:DCYCle?
Returns the duty cycle of the current trace data.
Unit: PCT
Usage: Query only
[SENSe<[1]..4>]:TRACe:MEASurement:PULSe:DURation?
Returns the pulse duration (pulse width) of the first detected pulse in the current trace.
Unit: S
Usage: Query only
[SENSe<[1]..4>]:TRACe:MEASurement:PULSe:PERiod?
Returns the time the pulse signal needs to complete one cycle.