Remote Control - Commands
NRP2
292User Manual 1173.9140.02 ─ 07
[SENSe<[1]..4>]:TRACe:MID:OFFSet:TIME
Defines an offset for pulse analysis on the time axis measured from the trigger event.
In conjunction with SENS:TRAC:MEAS:TIME this command can be used to define a
gate on a trace measurement, in which the pulse analysis shall be performed. If the
gate spans the same time interval as the trace measurement, the first detected pulse
will be analyzed. Using a gate, it is possible to investigate successive pulses.
Parameters:
<time> float_value
Range:
depending on the sensor
*
RST: depending on the sensor
Default unit:
S
[SENSe<[1]..4>]:TRACe:MID:TIME
Defines the considered length for pulse analysis on the time axis. In conjunction with
SENS:TRAC:MEAS:OFFS:TIME this command can be used to define a gate on a trace
measurement, in which the pulse analysis shall be performed. If the gate spans the
same time interval as the trace measurement, the first detected pulse will be analyzed.
Using a gate, it is possible to investigate successive pulses.
Parameters:
<time> float_value
Range: depending on the sensor
*
RST: depending on the sensor
Default unit:
S
[SENSe<[1]..4>]:TRACe:OFFSet:TIME <time>
Determines the relative position of the trigger event in relation to the beginning of the
Trace measurement sequence.
Parameters:
<time> float_value
Range: depending on sensor
*RST: depending on sensor
Default unit:
S
[SENSe<[1]..4>]:TRACe:POINts <points>
Sets the number of desired values per Trace sequence.
Error messages:
28 "Sensor not idle": The sensor is not in the IDLE state.