EasyManua.ls Logo

Rohde & Schwarz NRP2 - Page 297

Rohde & Schwarz NRP2
419 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control - Commands
NRP2
294User Manual 1173.9140.02 ─ 07
SERVice<[1]..4>:CALibration:TEST? <NR1>
Tests the consistency of the calibration data set according to three criteria:
If the following bit is set in
<nr1>,
the following action is executed.
0 The serial number of the calibration data set is tested.
1 The checksum of the calibration data set is tested.
2 The header of the calibration data set is tested.
If an inconsistency of the data set is detected, the query returns a 1, otherwise a 0.
Parameters:
<NR1> int_value
Range: 0 to 255
Usage: Query only
SERVice<[1]..4>:CALibration:TEMP <state>
Initiates a temperature measurement of the sensor in question. The temperature can
be queried with SERV:CAL:TEMP:DATA?.
Parameters:
<state> ONCE
Usage: Event
SERVice<[1]..4>:CALibration:TEMP:DATA?
Returns the temperature of the sensor in Kelvin if a temperature measurement was
previously initiated by means of SERV:CAL:TEMP
ONCE.
Usage: Query only
SERVice:DISPlay:OVERload <state>
Switches the sensor overload warning on/off.
Attention:
If the sensor is overloaded, it could be damaged! The warning should not be sup-
pressed if you are not certain that the sensor is out of danger.