Remote Control - Commands
NRP2
295User Manual 1173.9140.02 ─ 07
SERVice<[1]..4>:RCOunt <nr1>
SERVice<[1]..4>:RESult <nr1>
Allows simulated measured values to be specified for test purposes. SERV:RCO speci-
fies the number of SERV:RES
pairs to follow. The first SERV:RES command specifies
the measured value in watts, the second SERV:RES command determines how often
the measured value should be obtained. The measurement sequences defined in this
way are repeated until the sensor is disconnected from the R&S NRP2 and reconnec-
ted.
Parameters:
<nr1> int_value
*
RST: none. The simulation mode is not exited in the case
of *RST. To again obtain real measured values,
SERV:RCO 0 must be set.
Example: The command sequence
SERV2:RCO 2
SERV2:
RES 3
SERV2:RES 2
SERV2:RES 4
SERV2:RES 1
makes sensor 2 return the measured values 3 W, 3 W, 4 W (, 3
W, 3 W, 4W, ....) from now on.
SERVice<[1]..4>:SERRors <state>
The sensors signal an error to the R&S NRP2 if samples are lost while a signal is sam-
pled (sampling), as may be the case with high sampling rates. In such a case, the
measurement is less accurate. Since this error is signaled as an SCPI error, smooth
operation of a remote-control program may be disrupted. For this reason, sampling
errors can be suppressed by means of SERVice<[1]..4>:SERRors
OFF.
Parameters:
<state> ON | OFF
*RST: The R&S NRP2 BIOS setting is adopted.
SERVice<[1]..4>:UNLock <password>
Before other SERVice system commands are accepted, they must be enabled with
SERV:UNL 1234
.
Parameters:
<password> 1234
SERVice