6.13 STATus
The status registers of the R&S NRP2 can be read and configured with commands of
the STATus system.
The R&S NRP2 uses the following status registers:
Standard Event Status Register (specified in IEEE 488.2. Reading and configuring
with *ESR? and *ESE. This register is not handled by the STATus system.)
Device Status Register
Questionable Status Register
Questionable Power Status Register
Questionable Window Status Register
Questionable Calibration Status Register
Operation Status Register
Operation Calibrating Status Register
Operation Measuring Status Register
Operation Trigger Status Register
Operation Sense Status Register
Operation Lower Limit Fail Status Register
Operation Upper Limit Fail Status Register
Except for the Standard Event Status Register, each of these registers comprises five
subregisters with which the function of the register can be configured.
These subregisters are called
EVENt
CONDition
NTRansition
PTRansition
ENABle
The registers are 16 bits wide but only the lower 15 bits are used. This prevents prob-
lems with programs that cannot handle integers without sign.
The STATus system uses the following types of commands:
Queries
return a decimal value between 0 and 32767 (=2
15-1
).
Configuration commands set the ENABle, PTRansition, NTRansition registers of a status register
and thus determine the response of the register to status changes in
the R&S NRP2.
STATus