Bit No. Meaning
4 Window 4 Upper Limit Fail
The measured value returned by the fourth Calculate block exceeds the upper limit value.
5 Window 5 Upper Limit Fail
The measured value returned by the fifth Calculate block exceeds the upper limit value.
6 Window 6 Upper Limit Fail
The measured value supplied by the sixth Calculate block exceeds the upper limit value.
7 Window 7 Upper Limit Fail
The measured value supplied by the seventh Calculate block exceeds the upper limit value.
8 Window 8 Upper Limit Fail
The measured value supplied by the eighth Calculate block exceeds the upper limit value.
9 to14 Not used
15 Bit 15 will never be used.
6.13.17 Questionable Power Status Register
The CONDition register contains information as to whether the measured power values
are questionable.
The Questionable Power Status Register can be read by the following commands:
STATus:
QUEStionable:
POWer[:SUMMary]:CONDition?
STATus:QUEStionable:POWer[:SUMMary][:EVENt]?
Table 6-29: Meaning of bits used in the Questionable Power Status Register
Bit No. Meaning
0 Not used
1 Sensor A Power
The measurement data of sensor A is corrupt.
2 Sensor B Power
The measurement data of sensor B is corrupt.
3 Sensor C Power
The measurement data of sensor C is corrupt.
4 Sensor D Power
The measurement data of sensor D is corrupt.
5 Sensor A please zero
The zero correction for sensor A is no longer correct and should be repeated.
6 Sensor B please zero
The zero correction for sensor B is no longer correct and should be repeated.
7 Sensor C please zero
The zero correction for sensor C is no longer correct and should be repeated.