Remote Control - Commands
NRP2
318User Manual 1173.9140.02 ─ 07
Table 6-32: Initialization of device status
Result Power On DCL, SDC
(Device
Clear,
Selected
Device
Clear)
*RST or
SYSTem:
PRESet
STATus:
PRESet
*CLS
Power-On-
Status
Clear
(*PSC)
Effect 0 1
Clear STB (Status Byte) and
ESR (Standard Event Status Register)
-- yes -- -- -- yes
Clear SRE (Service Request Enable)
and
ESE (Standard Event Status Enable)
-- yes -- -- -- --
Clear PPE (Parallel Poll Enable) -- yes -- -- -- --
Clear EVENt register -- yes -- -- -- yes
Clear ENABle register of all OPERation
and
QUEStionable registers
-- yes -- -- yes --
Fill PTRansition register with "1",
delete NTRansition register
-- yes -- -- yes --
Clear error/event queue yes yes -- -- -- yes
Clear output queue yes yes yes 1) 1) 1)
Stop command processing (parser) and
clear input queue
yes yes yes -- -- --
1): Any command sent to a non-empty output queue causes the error:
-410 "Query interrupted"
6.14 SYSTem
With the aid of the SYSTem command system, administrative device settings can be
made and queried.
This includes:
●
list of error messages
●
IEC/IEEE-bus address
●
resetting the R&S NRP2
●
setting the date and time
●
audible signals
●
version numbers of hardware and software
●
setting the system speed
SYSTem