Remote Control - Commands
NRP2
320User Manual 1173.9140.02 ─ 07
Usage: Event
SYSTem:BEEPer:NOTify:KEY <state>
Enables (ON) or disables (OFF) acoustic key clicks, respectively.
Parameters:
<state> ON | OFF
*
RST: OFF
SYSTem:BEEPer:STOP
Stops the output of an audible signal.
Usage: Event
SYSTem:BEEPer:TIME <nrf>
Sets the length in milliseconds of the audible signal output with SYS:BEEP.
Parameters:
<nrf> int_value
Range: 1...60
*RST: This setting is not changed by *RST.
Default unit:
S
SYSTem:COMMunicate:GPIB[:SELF]:ADDRess <nr1>
Sets the address with which the R&S NRP2 can be addressed via the IEC/IEEE bus.
The address is factory-set to 20 and is not changed by a reset.
Parameters:
<nr1> int_value
Range: 0...30
*RST: This setting is not changed by *RST.
Usage: Event
SYSTem:COMMunicate:INET:MODE <mode>
Selects the mode for transmitting the special control sequences via ethernet according
to draft IEEE 1174.
Parameter Mode
0 no control sequences (abbr.: 1174.0 mode)
1 the IEC/IEEE 488.1 mode (abbr.: 1174.1 mode)
2 the IEC/IEEE 488.2 mode (abbr.: 1174.2 mode)