330User Manual 1173.9140.02 ─ 07
INITiate<[1]..4>:DISable <state>
INITiate:ALL:DISable <state>
This command prevents the execution of the INIT:IMM
command for one or more
sensors. It can thus suppress the start of a measurement for specific sensors if the
INIT:ALL:IMM command is used.
Parameters:
<state> ON | OFF
*RST: OFF
INITiate<[1]..4>[:IMMediate]
INITiate:ALL[:IMMediate]
This command starts a single-shot measurement. The respective sensor goes to the
INITIATED state. The command is completely executed when the sensor returns to the
IDLE state. The command is ignored when the sensor is not in the IDLE state or when
continuous measurements are selected (INIT:CONT ON
). The command is only fully
executed when the measurement is completed and the trigger system has again
reached the IDLE state. Besides CAL:ZERO:AUTO INIT is the only remote control
command that permits overlapping execution. Other commands can be received and
processed while the command is being executed.
Note:
INIT:IMM invalidates all previous measuring results. A FETCh? command following
INIT will thus always return a new measurement result.
Usage: Event
Error messages:
28 "Sensor not idle": The sensor is not in the IDLE state.
TRIGger:ALL:SYNChronize[:STATe] <state>
Using this command one can assure, that measurements with externally triggered sen-
sors start synchronously. As the initiation of measurements in the sensors is serialized
in the R&S NRP2 it is possible that a sensor is triggered before another sensor is initi-
ated (if synchronization is off). Sensors to be synchronized are required to have the
same TRIG:SLOP
setting, otherwise a SCPI error -221, "Settings conflict" occurs on
TRIG:ALL:SYNC ON.
Note:
TRIG:ALL:SYNC works only in conjunction with INIT:ALL.
Parameters:
<state> ON | OFF
*RST: OFF
TRIGger